Standards of Irradiation TID Tests for Space Applications
- Posted by José Joaquín González
- On October 30, 2020
- 0
Focus in most significative features of irradiation testing according to different standards, is essential after determine if it is necessary to perform an irradiation test. (link al otro artículo)
This question is one of the most common problems which we have to deal with when we want to perform an irradiation test (more specific a Total Ionizing Dose – TID – test) is how different is one irradiation standard from another. Are we accomplish the particular point of this or that standard when we refer to one specific aspect of our test once the irradiation standard was chosen.
Alter Technology has accumulated significant experience, conducting more than 1000 radiation test campaigns over more than 27 years for ESA projects, and commercial Space Programmes for Space agencies worldwide.
MIL-STD-883 TM1019.9
1.1 Irradiation Sources
This method describes the use of gamma rays using a Cobalt 60 source only.
1.2 Dose rates
Five conditions are expressed in this MIL-STD method depending on considered dose rate range or, in some situation, a variety of dose rate is not specified. Still, a condition of accelerated annealing is referred.
The next table summarizes these aspects:
MIL-STD-883 TM1019.9
wdt_ID | CONDITION | DOSE RATE RANGE |
---|---|---|
1 | A (Standard Condition) | 180 - 1080 krad(si)/h |
4 | B (Only for MOS Devices) | >180krad(Si)/h (if the dose rate in the intended application is lower than 180krad (Si)/h it can be accordated between the parties) |
7 | C (AS ALTERNATIVE) | WHATEVER DOSE RATE IF AGREEMENT BETWEEN PARTIES |
10 | D (ONLY BIPOLAR, CMOS or MIXED-SIGNAL) | |
13 | E (ONLY BIPOLAR, CMOS or MIXED-SIGNAL) | ACCELERATED TEST (NOT A DOSE RATE RANGE) |
CONDITION | DOSE RATE RANGE |
1.3 Irradiation room temperature range
For this standard, the temperature range is established in the interval: 18ºC≤IRRADIATION ROOM TEMPERATURE≤30ºC.
1.4 Electrical measurement temperature range
For this feature is specified the same range than in the previous irradiation temperature.
1.5 Irradiation temperature transport
In the case of the remote test (in a location different than the irradiation room), the samples must remain at a temperature lower than 10ºC plus the irradiation temperature. During the transport, all leads must be shorted.
1.6 Time to give back the samples to the irradiation area between irradiation steps.
This method establishes the samples must be replaced in the same irradiation position between two irradiation steps before two hours from stopping.
2 MIL-STD-750 TM1019.5
2.1 Irradiation Sources
Same the previous method, in this case, the use of gamma rays using a Cobalt 60 source is mandatory.
2.2 Dose rates
Three conditions are expressed in this MIL-STD method depending on considered dose rate range or, in some condition, a range of dose rate is not specified. Still, a condition of accelerated annealing is referred.
The next table summarizes these aspects:
MIL-STD-750 TM1019.5
wdt_ID | CONDITION | DOSE RATE RANGE |
---|---|---|
1 | A | 180-720 krad(Si)/h |
4 | B (ONLY FOR MOS DEVICES) | If the maxium dose rate is less than |
7 | C (AS ALTERNATIVE) | The test may be perfomed at the dose rate of the intended application, if agreement |
2.3 Irradiation room temperature range
For this standard the temperature range is established in the interval: 18ºC ≤ IRRADIATION ROOM TEMPERATURE ≤ 30ºC.
2.4 Electrical measurement temperature range
The range is established within the Interval: 20ºC ≤ ROOM TEMPERATURE ≤30ºC.
2.5 Irradiation temperature transport
In the case of the remote test (in a location different than the irradiation room), the samples must remain at a temperature lower than 10ºC plus the irradiation temperature. During the transport, all leads must be shorted.
2.6 Time to give back the samples to the irradiation area between irradiation steps.
This method establishes the samples must be replaced in the same irradiation position between two irradiation steps before two hours from stopping and, besides, it is established in one hour the time from the extraction and start of the electrical measurements.
3 ESCC22900 Issue 5
3.1 Irradiation Sources
This method allows using either cobalt 60 gamma-ray sources or an electron source.
3.2 Dose rates
Two conditions are expressed in this ESCC standard depending on the considered dose rate range.
The next table summarizes these aspects:
ESSCC22900 Issue 5
wdt_ID | CONDITION | DOSE RATE RANGE |
---|---|---|
1 | STANDARD RATE | 360-180000 rad(Si)/h |
4 | LOW RATE | 36-360 rad(Si)/h |
It is possible to study ELDR (Enhance Low Dose Rate) in the case of evaluation for devices that contains bipolar and optoelectronics within a low dose rate window for discrete devices or within one range of dose rate lower than 36rad(Si)/h for integrated circuits.
It is possible to study ELDR (Enhance Low Dose Rate) in the case of evaluation for devices that contains bipolar and optoelectronics within a low dose rate window for discrete devices or within one range of dose rate lower than 36rad(Si)/h for integrated circuits.
3.3 Irradiation room temperature range
For this standard the temperature range is established in the interval: 10ºC≤ IRRADIATION ROOM TEMPERATURE ≤30ºC. Variations of more than 3ºC are not allowed.
3.4 Electrical measurement temperature range
There is no temperature range for this point in this standard. The standards refer to use the temperature indicated in the detail specification.
3.5 Irradiation temperature transport
In the case of the remote test (in a location different than the irradiation room), the samples must remain at a temperature lower than 10ºC plus the irradiation temperature. During the transport, all leads must be shorted.
3.6 Time to give back the samples to the irradiation area between irradiation steps.
This method establishes as in the previous one; the samples must be replaced in the same irradiation position between two irradiation steps before two hours from stopping and, besides, it is established in one hour the time from the extraction and start of the electrical measurements.
4 ASTMF 1892-12 (2018)
4.1 Irradiation Sources
This method allows using either cobalt 60 or Cesium 137gamma ray source or X-Ray source, if low energy (lower than 10keV).
4.2 Dose rates
Four conditions are expressed in this ASTM standard depending on the considered dose rate range.
It is important to highlight that this standard considers the target material as Si2O instead of Si as in the previous standards referred to in this article.
The next table summarizes the aspects related to dose rate range:
ASTMF 1892-12
wdt_ID | CONDITION | DOSE RATE RANGE |
---|---|---|
1 | A (STANDARD DOSE RATE) | 180KRAD(Si2O)/h < DR < 1080krad(Si2O)/h |
4 | B (AS ALTERNATIVE) | The test may be performed at the dose rate of the intented application, if agreement. |
7 | C | If the maximum dose rate is less than |
10 | D (ONLY BIPOLAR, CMOS or MIXED-SIGNAL) | To meet unusual requirement and if agreed upon between parties to the thes, a dose rate that fits none of previous conditions may be used |
4.3 Irradiation room temperature range
For this standard, the temperature range is established in the interval: 18ºC ≤ IRRADIATION ROOM TEMPERATURE ≤ 30ºC.
4.4 Electrical measurement temperature range
For this feature is specified the same range than in the previous irradiation temperature criteria.
4.5 Irradiation temperature transport
In the case of the remote test (in a location different than the irradiation room), the samples must remain at a temperature lower than 10ºC plus the irradiation temperature. During the transport, all leads must be shorted.
4.6 Time to give back the samples to the irradiation area between irradiation steps.
This method establishes the time must be minimized and recorded. When CONDITION A of dose rate is used, this time between irradiation and test must be <10% of the incremental irradiation time (or 1h, whichever is greater) and the time between irradiations maybe 20% of incremental irradiation time (or 2h, whichever is greater).
- Standards of Irradiation TID Tests for Space Applications - October 30, 2020
- Radiation Effects and Test Flow for TID Testing - October 29, 2020
0 comments on Standards of Irradiation TID Tests for Space Applications